Online Raman Analyzer for Gases

Short description

Capable of detecting all gases except noble gases, enables simultaneous online analysis of multiple gas components, with a detection range from ppm to 100%.

RS2600-800800

Technical highlights

• Multi-component: simultaneous online analysis of multiple gases.
• Universal: 500+ gases can be measured, including symmetric molecules (N2, H2, F2, Cl2, etc.), and gas isotopologues (H2, D2,T2, etc.).
• Rapid response: < 2 seconds.
• Maintenance-free: can withstand high pressure, direct detection without consumables (no chromatographic column or carrier gas).
• Wide quantitative range: ppm ~ 100%.

Introduction

Based on Raman spectroscopy, Raman gas analyzer can detect all gases except noble gases (He, Ne, Ar, Kr, Xe, Rn, Og), and can realize the simultaneous online analysis of multi-component gases.

Following gases can be measured:

•  CH4, C2H6, C3H8, C2H4 and other hydocarbon gases in petrochemical field

•  F2, BF3, PF5, SF6, HCl, HF and other corrosive gases in fluorine chemical industry and electronic gas industry

•  N2, H2, O2, CO2, CO, etc. in the metallurgical industry

•  HN3, H2S, O2, CO2, and other fermentation gas in the pharmaceutical industry

•  Gas isotopologues including H2, D2, T2, HD, HT, DT

•  ...

de056874d94b75952345646937ada0d

Software Functions

The gas analyzer adopts the quantitative model of multiple standard curves, combined with the chemometric method, to establish the relationship between the spectral signal (peak intensity or peak area) and the content of multi-component substances.

Changes in sample gas pressure and test conditions do not affect the accuracy of quantitative results, and there is no need to establish a separate quantitative model for each component.

RS2600-front800800

Usage/implementation

Through valve control, it can achieve the functions of reaction monitoring:

• Monitoring the concentration of each component in reactant gas.
• Alarm for impurities in reactant gas.
• Monitoring the concentration of each component in exhaust gas.
• Alarm for hazardous gases in exhaust gas.

 

Typical applications

1709866044375
1709866161401
1709866197722
1709955647550
1709866320048
1709866371743

Product specifications

Product Model RS2600
 

 

 

 

 

 

Design/Appearance

RS2600-front800800
Air interface Standard ferrules, 3 mm, 6 mm, 1/8", 1/4" available
Connection interface USB2.0, RS232 DB9, RJ45
Pre-heating time <10 min
Power supply 100 ~ 240 VAC ,50 ~ 60 Hz
Sample gas temperature 30 ℃ ~ 40 ℃
Sample gas pressure <1.0 MPa
Working temperature 5 ℃ ~ 40 ℃
Humidity 0 ~ 60% RH
Screen 10 inch touch screen
Diemension 485 mm(width)× 350 mm(height)× 600 m(depth)
Weight 40 kg
Detecting components Hydrocarbons such as CH4, C2H6, C3H8, H2, DMK, etc.

Corrosive gases such as PF5, HCl, HF, POF3, etc.

Atmospheric components such as N2, O2, CO2, CO, H2S, etc.

Isotopic gases such as H2, D2, T2, HD, HT, DT, etc.